It was reported that implant #30 was placed on (b)(6) 2026, along with implants #29 and #27.On (b)(6) 2026, the patient presented for a final check and x-ray, which revealed bone loss and infection around implant #30.The patient returned on (b)(6) 2026 for the removal of implant #30.The area was bone grafted again, and a re-do of implant #30 will be attempted in 3+ months.No impact on the patient reported.
|
|
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided
a summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|