| Model Number |
TFL-FBX150BS |
| Medical Device Problem Codes |
Break (1069); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the single use fiber was broken by the end that plugs into the soltive system.The user fired the laser system without aiming beam present and a small flame was observed and was subsequently extinguished by the scrub nurse.The issue occurred at the beginning of the laser lithotripsy procedure that was completed with a similar device.There was no patient or user harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|