| Model Number |
CF-H190L |
| Medical Device Problem Codes |
Poor Quality Image (1408); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope, which is an olympus asset with no reported complaint.During the evaluation of the device repair technician found foreign material on the air water tube and air water cylinder, and noise image occurs.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device initial findings malfunction was confirmed.The initial evaluation findings were confirmed.Based on the final results of the investigation, the most probable cause of the foreign material on the air water tube and air water cylinder was not established.Also, for the noise image was due to damage on plug unit.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|