A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 23-dec-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device's inventory data not updating correctly across multiple pyxis stations.A technical support specialist (tss) verified communication between the stations and the server, confirmed there were no upload or download issues, and validated the station time settings.The tss then remotely accessed the server and resent count inventory and load messages for all loaded storage spaces to the affected stations.After the messages were processed, the customer confirmed that the inventory counts were displaying correctly and that the issue was resolved.The system functioned as intended after the technical support specialist troubleshot the device.
|