| Model Number |
FX814T |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
| Date of Event |
05/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an m.Blue shuntsystem (#fx816t) was implanted on (b)(6) 2024.According to the complainant, the shunt system caused an underdrainage.The patient underwent a revision procedure on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Permeability test:
a permeability test has shown that all components are permeable.Computer controlled test:
to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in vertical/ or both the horizontal as well as the vertical positions.The results show that the m.Blue operates within the accepted tolerance in both positions.Adjustment test:
the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test:
the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection:
after dismantling of the valve, no organic deposits were found.Results:
based on the investigation results, no functional deviations or other abnormalities were found in the product.The valve is operating within specifications.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
| |
|