| Model Number |
0998-00-0800-65 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 - event site postal code: h3t 1e2.\
\
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cardiosave intra-aortic balloon pump (iabp) helium gas loss message occurred.There was patient involvement.No harm was reported.It was reported that pump was replaced, the clinical staff kept the same balloon and replaced the cardiosave unit.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: h2, h11, g6, g3, b5, h6 (type of investigation, component codes), d9, h3\
upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer that during use, the cardiosave intra-aortic balloon pump (iabp) helium gas loss message occurred.There was patient involvement.No harm was reported.It was reported that pump was replaced, the clinical staff kept the same balloon and replaced the cardiosave unit.The getinge fst was dispatched to evaluate the unit.It was reported that no helium gas loss or fault nr 53 were observed.The software version was reinstalled as a precaution.Pim module, safety disk, power management board and solenoid board were replaced.The cardiosave device has been successfully maintained.It is now ready for clinical use.
|
| |
|