Ethicon Endo-Surgery, LLC. LIGACLIP; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown surgical procedure, the er320 clip applicator, lot 583d61, presented a functional failure at the time of use, since it did not trigger any clip.The event was identified during the procedure and the product was withdrawn for evaluation.No harm to the patient was reported.Was there any harm to the reported patient or user? no.Was there a significant delay? no.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/14/2026.D4 batch # z70515.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er320 device was returned with a clip in jaws.The clip was removed in order to inspect the jaws and they were found with no damage.In an attempt to replicate the reported incident, the device was tested for functionality; however, the trigger could not be actuated due to a jammed condition.Upon inspection, both the upper and lower forks of the feeder plate were found lodged within the track.The adjacent areas of the track were intact and showed no evidence of damage, deformation, wear, or other abnormalities.The device was subsequently disassembled to evaluate the condition of the internal components.Internal examination revealed that the trigger was bent, preventing proper advancement of the clips into the jaws.Additionally, twelve (12) clips were found remaining within the clip track at the time of disassembly.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.A manufacturing record evaluation was performed for the finished device batch z70515 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
PR  
+1(303)552-6892
MDR Report Key50752459
Report Number3005075853-2026-07490
Device Sequence Number18451944
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public(01)10705036012580(17)300430(10)583D61
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K812291
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberER320
Device Lot Number583D61
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/01/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-