Catalog Number DYND19124
Medical Device Problem Codes Disconnection (1171); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the tubing keeps slipping off the stop cock and water is spraying out while the surgeon is using it.".
 
Additional Manufacturer Narrative
The customer reported issues with the cysto tubing (4 lead) that comes in custom packs from cardinal; the tubing keeps slipping off the stop cock and water is spraying out while the surgeon is using it, and the 4 lead rubber is really soft.Urology qa launched testing at the dvv lab, confirmed failures in the associated lots from customer complaints, and included the 4-lead version of the product in testing.As this is a known defect, documented by both the vendor and urology qa, this complaint and any others describing this issue will be confirmed; the tubing issue was the result of a compromised batch of raw material from the vendor's supplier.The customer reported no patient was involved, the surgery was able to be completed, and no serious injury or medical intervention/attention was required.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SET,IRRIGATION,4-LEAD,90
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
shravan ingole
three lakes drive
northfield, IL 60093--275
MDR Report Key50752559
Report Number1417592-2026-38967
Device Sequence Number9203171
Product Code LJH
UDI-Device Identifier40080196993834
UDI-Public40080196993834
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K970946
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND19124
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/21/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2023
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1