The customer reported that one of the girls from the operating room pulled this pack from the case and the pack was not sealed; per rep: the white tyvek seal.A photo was received for the investigation, and the photo shows one unopen pack sitting outside and on top of the breather pouch; finished good production records and finished good pack build were reviewed, and there were no issues that might have had an impact on this issue.With the photo provided, medline was unable to confirm the issue and determine the root cause.The complaint was provided to the vendor for further evaluation, and the supplier quality team will also monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|