A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 11-sep-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the patients were not shown in the pyxis. a technical support specialist (tss) connected to the server and reviewed site down queries, service logs, iis application pools, and interface communications.Found that no admit discharge transfer messages had been received within the retention period and confirmed the cce was not receiving messages from the customer's adt host.Restarted the services and verified the interface reconnected, but no issues were identified within the pyxis environment.Worked with iest, who confirmed the issue originated from the customer¿s hit/adt system.The customer later confirmed that patients were appearing in pyxis normally, and the case was closed. the system functioned as intended after the technical support specialist troubleshot the device.
|