The account reported finding an adhesive issue with the drapes; patient contact was reported, and patient was not harmed or injured, with no patient infections or injury noted and it would just come unstuck.A sample of a used drape was received for investigation, finished good production records were reviewed, the finished good pack build was reviewed, and trending was reviewed.With the received sample medline was able to confirm the issue; the sample was sticking to itself and appeared to have been used, and there was no other physical damage.Medline was unable to determine the root cause with the received sample; this complaint was provided to the vendor for further evaluation, no response was received from the vendor after two follow ups, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|