The customer reported that "the cuff is leaking air and can pump it up, but the edge is leaking." medline received the cuff of a wall-mount aneroid blood pressure monitor; visual inspection found no cosmetic damage visible, and functional testing attached the monitor and bulb and filled the cuff with air, but it did not fill past 30mmhg.The complaint was confirmed, and the cause was documented as "comp./part defective, broke." no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|