Catalog Number MDS9400
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cuff is leaking air and can pump it up but the edge is leaking.
 
Additional Manufacturer Narrative
The customer reported that "the cuff is leaking air and can pump it up, but the edge is leaking." medline received the cuff of a wall-mount aneroid blood pressure monitor; visual inspection found no cosmetic damage visible, and functional testing attached the monitor and bulb and filled the cuff with air, but it did not fill past 30mmhg.The complaint was confirmed, and the cause was documented as "comp./part defective, broke." no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANEROID,WALL MOUNT,ADULT CUFF
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
madelyn ruppert
three lakes drive
northfield, IL 60093--275
MDR Report Key50753341
Report Number1417592-2026-39066
Device Sequence Number9460433
Product Code DXQ
UDI-Device Identifier30080196771268
UDI-Public30080196771268
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K211084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2024
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1