Model Number DTMA1QQ
Medical Device Problem Codes Electromagnetic Interference (1194); Over-Sensing (1438); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a device nurse indicated that the patient experienced electrical interference while using a meat grinder that was not properly grounded, leading to concerns about oversensing on the cardiac resynchronization therapy defibrillator (crt-d) ventricular lead.  information received indicted that due to electromagnetic interference (emi) a lead integrity alert (lia) triggered for high rate-non sustained episodes, sensing integrity counter (sic) (short v-v intervals) and rv bipolar lead impedance.  the crt-d remains in use.  it was also reported that the right atrial (ra) lead exhibited atrial lead position check failed.  the ra lead remains in use.  no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 6947m62 lead implanted: (b)(6) 2019 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Claria MRI Quad CRT-D SureScan
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50753480
Report Number3004209178-2026-15486
Device Sequence Number9414807
Product Code NIK
UDI-Device Identifier00643169720633
UDI-Public(01)00643169720633(17)190614
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/14/2019
Device Model NumberDTMA1QQ
Device Catalogue NumberDTMA1QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/14/2026
Date Device Manufactured12/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-45 LEAD, 429888 LEAD
Patient Age60 YR
Patient SexMale
Patient Weight236 LB
-
-