Discovered : during clinical procedure.Provide detailed description of problem : it was reported that the patient presented for implant uncovery, at which time approximately 50% bone loss was discovered around the implant at tooth site 30.The implant was removed, the site was curetted, and a bone graft was placed.Patient impact : no harm to patient.
|
|
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.D9: device availability unknown / not provided.G4: premarket identification k013227.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|