| Model Number |
ENF-VH |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the rhino laryngo videoscope to the service center for a separate issue.During the initial device analysis, the olympus service repair technician indicated that the insertion tube coating was peeling off.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: stress problem identified.The most probable cause of the stress problem was due to a cause traced to component failure, which is an expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|