| Model Number |
VH-4020 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Event site name exceeded character limitations: (b)(6).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
| |
|
Event or Problem Description
|
|
The hospital reported prior to use that there is an issue with the hemopro2 adaptor.The device does not supply enough power for cauterization.No risk to patients.The hemopro2 adaptor between the pliers and the power supply had a loose connection.This caused the caulking pliers to lose power.The cable was replaced by the customer.
|
| |
|