• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Defibtech, L.L.C. ReviveR; AUTOMATED EXTERNAL DEFIBRILLATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Defibtech, L.L.C. ReviveR; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DDU-100B
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that their aed's asi is flashing red, and the aed will not power on.The customer did not report this occurring during patient use.
 
Additional Manufacturer Narrative
The customer reported that the aed would not power on and that the aed's asi was flashing red.A flashing red asi indicates that the aed is off and that the aed or battery pack requires attention.Per the instructions for use, if the asi is flashing red and/or the unit emits a periodic beep, the unit may require servicing, the battery pack may be nonfunctional, or the defibrillation pads may not be connected to the device.There was no indication that the aed was stored without the pads attached.The aed was unable to complete a power on self-test during troubleshooting with customer service.Because a spare battery pack was not available, no further troubleshooting could be performed, and the device log files could not be retrieved.The customer was referred to an authorized distributor to purchase a replacement battery pack.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ReviveR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
Defibtech, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
Defibtech, L.L.C.
14 commercial st.
branford
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
+1(203)453-6654
MDR Report Key50753854
Report Number3003521780-2026-01557
Device Sequence Number21240432
Product Code MKJ
UDI-Device Identifier00815098020027
UDI-Public(01)00815098020027
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-100B
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-