| Model Number |
DDU-120E |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An international distributor reported that their customer's aed was reporting a battery low warning.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Although the customer initially reported a battery low warning, review of the aed's log files found the battery went on to fully deplete within the unit.On (b)(6) 2026, the aed identified that the aed battery was getting low and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until on (b)(6) 2026 when the battery pack became completely depleted.Analysis of the aed's log files did not identify a cause for the depleted battery.Analysis of the aed's log files identified that once a new battery was installed and a manual self test run the aed was functioning as designed and could stay in service.
|
| |
|