• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Defibtech, L.L.C. Lifeline; AUTOMATED EXTERNAL DEFIBRILLATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Defibtech, L.L.C. Lifeline; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DDU-120E
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
An international distributor reported that their customer's aed would not power on; however, when tested with a spare battery pack, it powered on.The customer did not report this occurring during patient use.
 
Additional Manufacturer Narrative
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.On (b)(6) 2025 when a replacement battery was installed the aed identified a previously uncleared service code and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until (b)(6) 2025 when the battery pack was measured at a critically low state and the aed began reporting replace battery pack now.The aed continued to flash and chirp until (b)(6) 2025 when the battery pack became completely depleted.There was no evidence in the electronic logs of any human interaction to address the aeds condition until (b)(6) 2026 when a replacement battery pack was inserted into the aed.The review identified that the aed functioned as designed and can stay in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Lifeline
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
Defibtech, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
Defibtech, L.L.C.
14 commercial st.
branford
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
+1(203)453-6654
MDR Report Key50753883
Report Number3003521780-2026-01572
Device Sequence Number9550941
Product Code MKJ
UDI-Device Identifier00815098020072
UDI-Public(01)00815098020072
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Distributor
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-120E
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-