This is report 1 of 2 of (b)(4).It was reported that during meniscal repair surgery, the meniscal deployment gun device would not release.The device would not release after being fired, resulting in the needle not being able to be removed from the meniscal gun device.The device was in use with an unknown implant.There was no issue with the needle.Another device was used to complete the procedure.There was no delay in the procedure reported.There were no reports of injuries, medical intervention or prolonged hospitalization. all available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the product has not returned to depuy synthes, however a photo was provided for evaluation.¿
visual inspection of the returned photo found the needle still attached to the meniscal gun.The overall complaint was confirmed as the observed condition of the meniscal deployment gun would have contributed to the complained device issue.¿
based on the findings, the investigation could not draw any conclusions and no potential cause about the reported event can be established due to the insufficient evidence provided.Multiple factors are associated with this type of failure, and the complaint device needs to be physically received and evaluated in our product analysis laboratory to determine a potential cause of why the customer experienced the failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.There was no non-conformance regarding this lot.
|