| Catalog Number |
2203E |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd smartsite vial access device had foreign matter in the fluid path.The following information was provided by the initial reporter: it was reported by customer that there was foreign matter.I am sharing a presentation where they are alleging a complaint(s) for vial adapters 2203e and 2203e-0006.
|
| |
|
Additional Manufacturer Narrative
|
|
B.3.The date of event is unknown; bd awareness date used.Device evaluation: no product or photo was returned by the customer.The customer complaint of foreign matter could not be verified due to the product not being returned for failure investigation.A device history record review for material# 2203e and lot# 25045202 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 09apr2025.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Search Alerts/Recalls
|