| Model Number |
M00584180 |
| Medical Device Problem Codes |
Break (1069); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a truetome 44 was attempted to be used in the papilla during an endoscopic retrograde cholangiopancreatography procedure to treat stones performed on (b)(6)2026.During the procedure, the cutting wire broke when performing sphincterotomy and the wire was lodged in the tissue.It was reported that a part of the cutting wire detached and fell into the patient and was eventually retrieved by pulling on the tome.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: the instructions for use (ifu) indicates that the sphincterotome does not need to be energized prior to performing sphincterotomy.Energizing the cutting wire prior to use will cause premature cutting wire fatigue and will compromise the cutting wire's integrity.However, the customer reported that the device was energized prior to bowing.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6:
imdrf device code a0401 captures the reportable event of cutting wire break.
|
| |
|