The customer reported issues with stick-on ekg leads, including ¿lots of noise, troubles sticking, and inconsistent reads.¿ complaint samples were checked for electrical performance test as per ansi/aami ec 12:2000, and no problem with signals was observed and values were within the controlled specifications; device history records and inspection/testing records were reviewed and found in compliance with specifications, and user test found the samples applied on dry skin didn¿t come off or detach from the skin.The complaint is not confirmed.No root cause was identified, and no trends were identified for the specified material number and corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|