| Model Number |
0293 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead triggered an alert for high, out of range shock impedance (lvsi) showing a gradual increase.It was discussed that the lead would require replacement when 30 day average becomes 150 ohms or greater.No adverse patient effects were reported.This rv lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|