| Model Number |
U225 |
| Medical Device Problem Codes |
Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited undersensing and a decrease in biventricular pacing.Technical services (ts) was consulted and confirmed the undersensing was through the non-boston scientific atrial lead.Ts informed the health care professional that this allows intrinsic beats to occur and prevents the biventricular pacing.Ts reviewed measurements and recommended programming atrial sensitivity to a nominal value.No adverse patient effects were reported.This crt-p remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|