It was reported that during shoulder clavicular joint dislocation surgery, when tightening the rigidloop suture device, the suture was broken off.Changed another one to continue the surgery, the same problem happened again.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.Report 2 of 2 for (b)(4).
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
the product has not returned to depuy synthes, however a photo was provided for review.Visual inspection of the returned photo found the white suture broken and frayed.No other anomalies were noted.The overall complaint was confirmed as the observed condition of the rgdloop adjustable stnd would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to use of snaps or other instrument to pull the suture without wrapping the suture around the snaps creates a stress point on the suture, resulting in breaking it.The white suture is used to assembly into a graft construction with provides fixation.The green / white suture is the lead suture due to it is used to pull the implant/ graft construction with tension, the tension values are associated with this final implant assembly.The tension clinical spec is high compared to expected intraoperative loads.Therefore, the tension is measured only on the green/white suture during the manufacturing process.Regarding the white suture, an inspection is performed during the manufacturing process to check the measurement value and the condition; this information correspond to a process control and it is the document to use to guarantee the inspection of the suture, due to this damage suture can¿t have happened during manufacturing process.As per the instructions for use, the steps for a proper insertion are provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review:
there was no non-conformance regarding this lot.
|