The customer reported that needles are popping off and not locking on, and noticed the needles were not locking on when they went to attached with prefilled syringe.Medline received 100ea of (b)(6) and tested 29ea randomly using a 3ml syringe and simulated blood draws using a phlebotomy training test set up.Of the 29ea tested none had the reported defect, and the investigation results were not confirmed.No trends were identified for the specified material number and its corresponding defect type, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|