Catalog Number DYNJRA1823
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "both of the stopcocks in the kits are leaking at a single juncture and appear to have hair line cracks.".
 
Additional Manufacturer Narrative
The account reported issues with the stopcocks in the tap block kits; both stopcocks were leaking at a single juncture, appeared to have hair-line cracks, and leakage increased medication waste during procedures.A sample was received for investigation consisting of five stopcocks attached to various size syringes; each syringe was filled with tap water to test the stopcocks for leaks, each stopcock failed due to a visible leak when the plunger was depressed, and cracks in the white valve that direct the fluid path were found.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TAP BLOCK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kim evangelista
three lakes drive
northfield, IL 60093--275
MDR Report Key50754574
Report Number1417592-2026-39250
Device Sequence Number21181118
Product Code OFT
UDI-Device Identifier40193489961882
UDI-Public40193489961882
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJRA1823
Device Lot Number22LMI568
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/22/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2023
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-