Catalog Number SYR100302
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 04/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the provider went to use the needle, the needle was not protected in the cap.The needle was sticking out past the cap causing the provider to be stuck by it.
 
Additional Manufacturer Narrative
According to the customer, on (b)(6) 2023 a provider was poked by a capped needle that was sticking out through the plastic cap; the stick was reported on the left thumb, the needle was clean, and the provider did not require any post exposure tasks.No samples were received for investigation, but a lot # was provided; reserve stock from the same lot was ordered, and 25 samples were randomly selected for a visual inspection and functional testing by attaching the needles to medline luer lock syringes and inserting the samples into a training pad for hub testing.The visual inspection revealed no manufacturing or any other defects preventing the samples to work as intended, all the needles remained covered by their safety cap within the sterile packaging, and no needles protruding from the cap or the cap detaching from the needle when attached to the hub was experienced.Based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications; no root cause was identified, and the samples worked as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
NEEDLE,30GX0.5"REG,BEVEL
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kehkashan khan
three lakes drive
northfield, IL 60093--275
MDR Report Key50754683
Report Number1417592-2026-39272
Device Sequence Number9148868
Product Code FMI
UDI-Device Identifier40080196305637
UDI-Public40080196305637
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K112057
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSYR100302
Device Lot Number220922
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2023
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-