| Model Number |
SC-8336-50 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
08/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's spinal cord stimulation (scs) lead had migrated to the left side of midline, while the patient required right-side coverage.Due to the patient's anatomy, the physician was unable to revise the lead to the desired location.As a result, the spinal cord stimulator (scs) system was explanted.All explanted device components will not be returned per facility policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event:
model number/catalog number: sc-1232
serial number: (b)(6).Batch/lot number: 820772
model/catalog description: wavewriter alpha 32 ipg kit
unique identifier (udi): (b)(4).Model number/catalog number: sc-4316
batch/lot number: 38610014
model/catalog description: clik anchor
unique identifier (udi): (b)(4).
|
| |
|