| Catalog Number |
07P72-20 |
| Medical Device Problem Code |
Imprecise or Erratic Results (4059)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported imprecise alinity c carbon dioxide (co2) results generated across multiple alinity c processing modules.The customer stated that the anion gaps were running low.The following data was provided (reference range: 22-29 mmol/l):
(b)(6) (b)(6): units: mmol/l
sid (b)(6): co2 initial = 30; repeat = 28
sid (b)(6): co2 initial = 30; repeat = 27
sid (b)(6): co2 initial = 30; repeat = 28
sid (b)(6): co2 initial = 31; repeat = 29
sid (b)(6): co2 initial = 22; repeat = 20.Thelma (b)(6): units: mmol/l
sid (b)(6): co2 initial = 30; repeat = 28
sid (b)(6): co2 initial = 30; repeat = 28
sid (b)(6): co2 initial = 31; repeat = 28
sid (b)(6): co2 initial = 22; repeat = 21.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section a patient information: refer to section b5 for complete patient information.
|
| |
|