• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-KNEE PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-KNEE PACK Back to Search Results
Catalog Number DYNJ46384N
Medical Device Problem Code Adhesive Too Strong (4071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The account reported finding coban adhesive issue, and the reporter stated the product was sticking and not pulling apart easily.A sample was received for investigation; medline received one coban from the customer¿s shoulder pack that was labeled good and one coban that was from the reported pack, reviewed finished good production records, reviewed the finished good pack build, and reviewed trending.Both cobans had no issues with adhesive and no other defects were noted; based on the information provided and with the physical sample received, medline could not confirm the reported issue.No visual defects were observed, medline was unable to confirm a potential root cause with the sample provided, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.This is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Event or Problem Description
It was reported that coban component (b)(6) in multiple packs it extremly difficult to get off of the roll.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-KNEE PACK
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
namirah khan
three lakes drive
northfield, IL 60093--275
MDR Report Key50754766
Report Number1417592-2026-39281
Device Sequence Number15759178
Product Code OJH
UDI-Device Identifier40193489937276
UDI-Public40193489937276
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNJ46384N
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/24/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2023
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-