A patient implanted with an evoke spinal cord stimulation (scs) system was evaluated for possible lead migration.The patient also reported discomfort at the evoke closed loop stimulator (cls) implant site due to contact with the patient¿s belt.X-ray imaging confirmed migration of both evoke leads.During the revision procedure, the physician noted that the anchors were functioning as intended and the patient¿s post-implant activity level may have contributed to the lead migration.One lead was replaced, the other lead was repositioned, the existing anchors were reused, the cls was repositioned to a higher location.
|
|
(b)(4)
cls information:
brand name: evoke closed loop stimulator (cls)
model: 102901
catalog: 3042
lot/batch number: 211e21
udi: (b)(4)
expiration date: 10april2027.The replaced lead was not returned to saluda.The root cause of the lead migration could not be definitively determined; however, the physician noted that the patient¿s activity level following implantation may have contributed to the lead migration.The evoke scs system surgical guide detail potential risks associated with surgery and spinal cord stimulation including, "lead migration or suboptimal placement, which may result in undesirable stimulation changes, and the patient may require surgery (including revision, explant, and replacement) as a result¿.The discomfort at the cls implant site was associated to the patient¿s belt line.The evoke scs system surgical guide details potential risks associated with surgery and spinal cord stimulation including, "temporary or persistent post-surgical pain at hardware implantation sites and the patient may require surgery (including revision, explant, and replacement) as result¿.
|