• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Saluda Medical PTY LTD Direct Clamp Active Anchor Kit; SCS Anchor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Saluda Medical PTY LTD Direct Clamp Active Anchor Kit; SCS Anchor Back to Search Results
Model Number 104523
Medical Device Problem Codes Patient-Device Incompatibility (2682); Migration (4003)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 08/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
A patient implanted with an evoke spinal cord stimulation (scs) system was evaluated for possible lead migration.The patient also reported discomfort at the evoke closed loop stimulator (cls) implant site due to contact with the patient¿s belt.X-ray imaging confirmed migration of both evoke leads.During the revision procedure, the physician noted that the anchors were functioning as intended and the patient¿s post-implant activity level may have contributed to the lead migration.One lead was replaced, the other lead was repositioned, the existing anchors were reused, the cls was repositioned to a higher location.
 
Additional Manufacturer Narrative
(b)(4) cls information: brand name: evoke closed loop stimulator (cls) model: 102901 catalog: 3042 lot/batch number: 211e21 udi: (b)(4) expiration date: 10april2027.The replaced lead was not returned to saluda.The root cause of the lead migration could not be definitively determined; however, the physician noted that the patient¿s activity level following implantation may have contributed to the lead migration.The evoke scs system surgical guide detail potential risks associated with surgery and spinal cord stimulation including, "lead migration or suboptimal placement, which may result in undesirable stimulation changes, and the patient may require surgery (including revision, explant, and replacement) as a result¿.The discomfort at the cls implant site was associated to the patient¿s belt line.The evoke scs system surgical guide details potential risks associated with surgery and spinal cord stimulation including, "temporary or persistent post-surgical pain at hardware implantation sites and the patient may require surgery (including revision, explant, and replacement) as result¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Direct Clamp Active Anchor Kit
Common Device Name
SCS Anchor
Manufacturer (Section D)
Saluda Medical PTY LTD
5 eden park drive
macquarie park, nsw 2113
AS  2113
Manufacturer (Section G)
Saluda Medical Pty Ltd
5 eden park drive
macquarie park, nsw 2113
AS   2113
Manufacturer Contact
sadie mauser
5 eden park drive
macquarie park, nsw 2113
AS   2113
MDR Report Key50754797
Report Number3021836309-2026-00276
Device Sequence Number9361611
Product Code LGW
UDI-Device Identifier09352307001817
UDI-Public09352307001817
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number104523
Device Catalogue Number3043
Device Lot Number9018427705
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-