| Catalog Number |
MX6760 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, after prep, the user was in the process of inserting the atraumatic tip of a 6/7f mynx control vascular closure device (vcd) into the sheath hub, when the user noticed fraying of the proximal tip of the sealant housing with exposed sealant.A new device from the same lot was prepared and deployed without complication.The dvice was removed from table and new mynx control vascular closure device 6f/7f was deployed without any issues or complications.There was no reported patient injury.The device was removed from packaging by handle and prepped according to the instructions for use (ifu) standards by a trained user.The device was inspected for damage when removed from the package, and no damage to the packaging or device was noted at that time.The device never entered the sheath.A6f non-cordis sheath was used.The device was never inserted into the patient or deployed.Femoral artery suitability, including a vascular sheath introducer insertion angle of 30¿45 degrees, was verified on angiography.The vessel diameter was verified to be greater than or equal to 5 mm.There was no peripheral vascular disease (pvd) or calcium in the vicinity of the puncture site.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.This device is available for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.
|
| |
|
Search Alerts/Recalls
|
|
|