• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; STRIP,CLOSURE,WOUND,MEDI-STRIP,1/4X3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; STRIP,CLOSURE,WOUND,MEDI-STRIP,1/4X3 Back to Search Results
Catalog Number NON250314
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Erythema (1840); Blister (4537)
Date of Event 12/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when applying this tape, it is causing blisters on patients.Also they are having issues with the tape not sticking.
 
Additional Manufacturer Narrative
The customer reported that when applying this tape, it is causing blisters on patients, and they are having issues with the tape not sticking.The provided photos were reviewed, the provided samples were evaluated, the vendor was notified of the issue and sent the provided samples, and trend analysis was performed by reviewing the past two years of complaint data.The reported issue was confirmed; however, medline could not confirm that the product failed to meet specifications, and no product defects were identified as a result of the sample evaluation.A review of the complaint information as well as the results of the investigation indicate no need for further action at this time, and a divisional quality representative will continue to monitor this issue for tracking and trending purposes during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
STRIP,CLOSURE,WOUND,MEDI-STRIP,1/4X3
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
daniel soto
three lakes drive
northfield, IL 60093--275
MDR Report Key50754865
Report Number1417592-2026-39309
Device Sequence Number9550437
Product Code KGX
UDI-Device Identifier20080196691859
UDI-Public20080196691859
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberNON250314
Device Lot Number471067
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/16/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2025
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-