• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; DBD-SAFETY BLOOD COLLECTION NEEDLE 22GX1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; DBD-SAFETY BLOOD COLLECTION NEEDLE 22GX1 Back to Search Results
Catalog Number DYNDSBCN22G
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the safety device on the needle is popping off during use.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline safety needle with the safety device popping off during use; according to the customer, during a normal venipuncture the patient had a painful experience from the stick due to the safety needle cover falling off during use, with no unintended needle stick outside of what was intended for the procedure and no serious injury, medical intervention, or follow-up care.Medline received one case of safety blood collection needles, randomly selected a total of 50 using samples from each box, used three vacutainer tubes with blood collection tube holders to simulate a multiple tube draw for each needle, then activated the safety locking mechanism and inspected for loose parts or inactivation of the safety device.All 50 needles passed with no parts falling off or misfiring of the safety device of any kind, all samples worked as intended, and the investigation results were not confirmed.Based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
DBD-SAFETY BLOOD COLLECTION NEEDLE 22GX1
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093
Manufacturer Contact
shivani gheewala
three lakes drive
northfield, IL 60093
MDR Report Key50755158
Report Number1417592-2026-39357
Device Sequence Number13178661
Product Code FMI
UDI-Device Identifier10193489007886
UDI-Public10193489007886
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182146
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDSBCN22G
Device Lot Number31824060001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/04/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2025
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-