• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; CATHETER,URETH,VINYL,UNIVERSAL,16",14 FR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; CATHETER,URETH,VINYL,UNIVERSAL,16",14 FR Back to Search Results
Catalog Number DYND10723H
Medical Device Problem Codes Unsealed Device Packaging (1444); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that (b)(4) came in damaged and it looks like a sealing issue.
 
Additional Manufacturer Narrative
The customer reported that (b)(4) came in damaged and it looks like a sealing issue.Medline quality received (b)(4) of the reported product in packaged, unused condition, and visual assessment of the provided product revealed that a long cut on the pouch was present, running along the packaging lengthwise; the catheter was exposed, and the sterile barrier was compromised.The reported issue of damaged packaging was verified by visual assessment of the provided sample, but the root cause could not be determined; the vendor completed the investigation form, and a review of their records, manufacturing records, and retained samples found no abnormalities or defects.There are currently no known trends for this product, and medline quality will continue to monitor for the issue through monthly complaint meetings and quarterly trending analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
CATHETER,URETH,VINYL,UNIVERSAL,16",14 FR
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
herchelle teodoro
three lakes drive
northfield, IL 60093--275
MDR Report Key50755302
Report Number1417592-2026-39396
Device Sequence Number21172546
Product Code EZD
UDI-Device Identifier10080196846825
UDI-Public10080196846825
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K091306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND10723H
Device Lot Number0332501A090
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/26/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2025
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-