| Model Number |
E141 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Low impedance (2285)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) system exhibited oversensing of noisy signals for its right ventricular (rv) lead.Technical services (ts) was consulted and reviewed stored data data.Ts discussed stored data, with a decrease in impedance measurements for the rv lead.The impedance measurements were still within range.Ts suspected a lead integrity issue based on the age of the rv lead.This device remains in service.No adverse patient effects were reported.
|
| |
|