According to the facility, staff members experienced strikethrough on their arms during a procedure; per the facility there was no impact to the staff member or patient, and no medical intervention, serious injury, or follow-up care was reported.A sample was received for investigation, three unused packs with three spt-2362p gowns were evaluated, finished good production records and finished good pack build were reviewed, and trending was reviewed.No damage or defect with the gown was observed; the gown has an aami level 3 rating, is fluid resistant but will not prevent strikethrough, and the item performed to specifications.The cause code was adjusted to dc: traced to user with knowledge, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|