• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. 4K CAMERA HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. 4K CAMERA HEAD Back to Search Results
Model Number CH-S400-XZ-EB
Medical Device Problem Codes Leak/Splash (1354); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported damage to the head during inspection for use involving an olympus camera head.During the initial evaluation of the returned device, compromised water tightness, was observed.There was no patient involvement.
 
Additional Manufacturer Narrative
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.A definitive root cause could not be identified.However, based on the results of the investigation, the most probable cause of compromised water tightness was attributed to a scratch on the cable unit.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50755633
Report Number3002808148-2026-32152
Device Sequence Number9550468
Product Code FET
UDI-Device Identifier04953170374609
UDI-Public04953170374609
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K172817
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCH-S400-XZ-EB
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-