| Catalog Number |
8065753075 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that during cataract surgery with intra ocular les implantation, an ophthalmic console failed to deliver the phacoemulsification power.There was some object/debris noted underneath the treadle.After the area was cleaned, the treadle position correctly reached phacoemulsification mode (position three) upon full depression, and the phacoemulsification power delivery issue was resolved.The surgery then started without any problems and was completed without any issues.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|