The account reported finding the breather pouch open.The investigation included a stock check, review of finished good production records, review of finished good pack build, trending review, and sample evaluation; the stock check found no defects, and the sample evaluation found the lower part of the tyvek portion was fully open.The reported problem was confirmed by the sample evaluation.The tyvek cover appeared wrinkled and tugged, indicating it is likely that the pack was mishandled and opened at some point after assembly; with the information provided, medline was not able to confirm at what point the damage occurred.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|