The customer reported that the curas x-large plastic adhesive bandages 2x4 ¿don¿t stick and keep falling off not even an hour after put on.¿ physical samples of non25504 were provided by the customers, and medline evaluated 6 months of complaint history for this product and had 1 similar complaint from a different vendor.Based on the samples provided, medline was unable to confirm the reported defect that this item has weak adhesion; the cause was not confirmed with received sample.Due to this reported issue being unconfirmed, the vendor will not be notified, a closing response has been disclosed, and the divisional quality representative will continue to monitor this issue for tracking and trending purposes during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|