• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medical Systems Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medical Systems Patient Information Center iX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number Patient Information Center iX
Medical Device Problem Code Intermittent Communication Failure (4038)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint on the patient information center ix indicating that the monitoring station was on pause even though two patients were being monitored.The device was in use on patient at time of event, there was no adverse event reported.
 
Additional Manufacturer Narrative
E1: (b)(6).E1: (b)(6).The service engineer evaluated the system after both monitors in room 2 were found in pause mode while two patients were being monitored; no error code was reported and there was no delay to the patient or procedure.The issue could not be reproduced.The pic ix central unit and x2 monitor logs were backed up and forwarded for further analysis, and a performance verification was completed.No device failure was confirmed; the cause remains unknown, and the issue did not recur after the monitor was returned to monitoring mode.No corrective action was taken, the customer reported no further abnormalities, and testing confirmed that the system met manufacturer specifications and was ready for clinical use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Patient Information Center iX
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
+1(617)245-5900
MDR Report Key50756258
Report Number1218950-2026-100628
Device Sequence Number9195317
Product Code MHX
UDI-Device Identifier00884838093041
UDI-Public(01)00884838093041(10)C.03.10
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K183387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPatient Information Center iX
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2026
Initial Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-