| Model Number |
Patient Information Center iX |
| Medical Device Problem Code |
Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on the patient information center ix indicating that the monitoring station was on pause even though two patients were being monitored.The device was in use on patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: (b)(6).E1: (b)(6).The service engineer evaluated the system after both monitors in room 2 were found in pause mode while two patients were being monitored; no error code was reported and there was no delay to the patient or procedure.The issue could not be reproduced.The pic ix central unit and x2 monitor logs were backed up and forwarded for further analysis, and a performance verification was completed.No device failure was confirmed; the cause remains unknown, and the issue did not recur after the monitor was returned to monitoring mode.No corrective action was taken, the customer reported no further abnormalities, and testing confirmed that the system met manufacturer specifications and was ready for clinical use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|