BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; Implantable lead Back to Search Results
Model Number 4543
Medical Device Problem Codes High impedance (1291); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this left ventricular (lv) lead exhibited high out of range pacing impedance measurements.It was later discovered this lv lead was fractured.The cause and location of the fracture on the lead are unknown.Subsequently, this lv lead was surgically abandoned and successfully replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Common Device Name
Implantable lead
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50756331
Report Number2124215-2026-48938
Device Sequence Number9395911
Product Code NIK
UDI-Device Identifier00802526410994
UDI-Public(01)00802526410994(17)160108
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/08/2016
Device Model Number4543
Device Catalogue Number4543
Device Lot Number187412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2026
Initial Report FDA Received Date09/14/2026
Date Device Manufactured01/08/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age87 YR
Patient SexMale
-
-