| Catalog Number |
MDT24SUPRADPS |
| Medical Device Problem Codes |
Pumping Problem (3016); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the lock button should activate after 3 minutes, it does not," "it alarms about a half hour/hour later," and "i have tried on different mattresses, but 3 pumps keep doing the same thing.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported that the lock button should activate after 3 minutes and it does not, and that it alarms about a half hour/hour later; the customer confirmed 3 pumps have the issue.The sample has been received and this pump was manufactured in nov of 2018; the warranty is 18 months and this is well outside warranty period.The investigation stated this was "not a confirmed complaint," and the updated cause code is "unknown responsible party with cannot determine cause w/ received sample." a review of the complaint information as well as the results of the investigation indicate no need for further action at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|
Search Alerts/Recalls
|
|
|