| Catalog Number |
103-1003 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision occurred on (b)(6) 2026 approximately after 14 days after the primary surgery which happened on (b)(6) 2026.No fall or trauma was reported.Based on comparing usage forms humeral cup stability pe/ta6v ø36/+3 and fx v135 humeral spacer ta6v +9mm was explanted due to cup and spacer disassociated from stem.This part was replaced with humeral cup standard pe/ta6v ø36/+3 and humeral spacer ta6v +9mm.Fx v135® humelock stem ta6v ø36/10 l.120mm cementless ti/hawas not replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Revision occurred approximately 14 days after the primary surgery due to disassembly.No actual, implied, or suspect link to fx product.
|
| |
|