• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices Back to Search Results
Model Number CDS0801-XTW
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) with a grade of 4 on a patient with a prolapsed posterior leaflet and pre-existing flail.A mitraclip xtw was inserted and deployed on the mitral valve.A second mitraclip xtw (60422r1004) was then prepared for use.However, while preparing the second mitraclip xtw, after testing the lock sequence, the clip would not open.Troubleshooting was performed but was not successful.Therefore, the clip was not used.The procedure was completed with one clip implanted, reducing mr to a grade of 1.There were no adverse patient effects and no clinically significant delay in the procedure.Device returned and analysis performed on (b)(6)2026 revealed that the mandrel was protruding from the proximal end of the actuator knob, the collet was broken, and there was a leak at the handle-arm positioner.
 
Additional Manufacturer Narrative
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MitraClip G5 Delivery System
Common Device Name
mitral valve repair devices
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50756712
Report Number2135147-2026-06202
Device Sequence Number13071704
Product Code NKM
UDI-Device Identifier05415067050779
UDI-Public(01)05415067050779(17)271021(10)60422R1004
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-XTW
Device Catalogue NumberCDS0801-XTW
Device Lot Number60422R1004
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/18/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Supplement Date Received by Manufacturer09/29/2026
Initial Report FDA Received Date09/14/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1 Implanted MitraClip; Steerable guide catheter
Patient SexUnknown
-
-