It was reported that the customer had an issue with the 30 degree endoscope scope movement, accompanied by noise from the mechanical parts.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the distributor field service engineer (fse) and obtained the following additional information: the endoscope moves slowly and produces an unusual noise during movement.In some cases, the endoscope would also rotate unexpectedly without user control.The customer confirmed that the endoscope moved freely with uncontrolled motion.There was no damage observed on the endoscope¿s adapter or base.Intuitive surgical, inc.(isi) followed up with another distributor field service engineer (fse) and obtained the following additional information: the fse clarified that as seen in the provided video, the scope had a mechanical noise, however, the endoscope worked without error with the surgeons.The endoscope worked normally with the surgeons until now.The fse heard that once there was a surgeon that felt a very brief uncontrolled movement.The fse did not have detailed information about this incident and it was not confirmed.The fse agreed with the hospital to exchange the scope and then the old scope would be sent back to intuitive for further analysis.There was no injury to the patient due to the issue.The procedure was completed with no delays.No fragment fell into the patient.
|