| Catalog Number |
FG-2000-00 |
| Medical Device Problem Code |
Incorrect Interpretation of Signal (1543)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during primary functional endoscopic sinus surgery (fess) to treat all sinuses, the accuracy of the trudi navigation system (fg-2000-00) was off by 6+ mm.Staff reported "accuracy showing us that we were in the orbit when that was not the case." it was reported that inaccuracy was determined by knowing where the suction and navigation devices were in the patient verses what the trudi was showing.When the issue was observed, the icon of the device appeared green on the trudi.It was reported that there was an extra seven (7) minutes in surgery time due to having to re-register the patient.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|